Parasternal Extravascular ICD System Pivotal Clinical Investigation (ALARION EV) Study
a study on Ventricular Arrythmia Ventricular Fibrillation Tachycardia Ventricular Tachycardia
Summary
- Eligibility
- for people ages 18 years and up (full criteria)
- Location
- at La Jolla, California and other locations
- Dates
- study startedstudy ends around
- Principal Investigator
- by Ulrika Birgersdotter-Green, MD
Description
Summary
The ALARION EV Study is a prospective, multi-center, single-arm, non-randomized study without concurrent or historical controls. The purpose of this pivotal study is to generate evidence of safety and effectiveness to apply for regulatory approvals for commercialization of the Parasternal EV-ICD System.
The objective of the study is to demonstrate safety and effectiveness of the Parasternal EV-ICD System through 6-months.
Details
The primary safety endpoint is Freedom from Major Adverse Device Effects (MADEs) through 6 months. The primary effectiveness endpoint is Defibrillation Performance: Successful induced ventricular arrhythmia (IVA) test in the final system configuration.
Up to 40 Investigational Sites Worldwide will participate with up to 280 subjects enrolled in the study. At least 50% of subjects will be enrolled at sites in the United States.
Subjects meeting indications for ICD therapy who do not require permanent bradycardia pacing will be eligible for participation. Eligible Subjects will receive an Atala™ lead connected to an EV-ICD Pulse Generator positioned in a left pectoral or left lateral location. The Parasternal EV-ICD System will be tested to verify appropriate sensing and defibrillation.
Starting at 6 months, Subjects will be followed every 6 months until the study is closed. The study is expected to be ongoing for a minimum of 2.5 years from the date of first enrollment. Endpoints will be analyzed and summarized in a final study report after all active Subjects have completed the 6-month visit.
Keywords
Ventricular Arrythmia, Ventricular Fibrillation, Ventricular Tachycardia, Ventricular Defibrillation, Extravascular, Atala™ lead and EV-ICD Pulse Generator, Parasternal EV-ICD System
Eligibility
You can join if…
Open to people ages 18 years and up
- At least 18 years old
- Class I or IIa indication for implantation of an ICD according to the ACC/AHA/HRS Guidelines or ESC guidelines
You CAN'T join if...
- Patients who cannot read or write
- Expected survival < 1 year
- Pregnant or nursing subjects and those who plan pregnancy during the clinical investigation follow-up period
- Participation in any concurrent clinical study without prior approval from the Sponsor
- Inability or unwillingness to provide informed consent to participate in the study
- Any known circumstances which may prevent the completion of protocol testing and data collection through the 6-month follow-up visit
- Circumstances that may prevent data collection or completion of specified follow-up visits
- Allergies to any device materials listed in the Instructions for Use (IFU)
- Contraindication for temporary suspension of oral/systemic anticoagulation
- Known history of lung disease with FEV1 < 1.0 Liter
Device Related:
- Implanted with or planned implantation of any device which delivers current in the body that may interfere with therapy delivery, including, but not limited to a pacemaker, LVAD or neurostimulator
- Implanted with a subcutaneous ICD lead, subcutaneous coils/arrays, epicardial patches or epicardial pace/sense leads
- Any known need for MRI prior to approval of MRI conditional labeling
Anatomy Related:
- Known structural abnormalities of the heart that may increase risk of the Parasternal EV-ICD System procedure
- Known obstructed or restricted pathway into the mediastinum for the Atala™ lead
- Prior surgery with disruption of the lung, pericardium or connective tissue between the sternum and pericardium, including sternotomy of any type
- Known significant anatomic derangement of or within the thorax (e.g., pectus excavatum, significant scoliosis)
- History of thoracic radiation therapy or other medical treatments/conditions which may complicate the Parasternal EV-ICD System implant procedure
- Known adhesions in the thorax or history of medical treatments, surgeries or conditions that increase the potential for adhesions in the thorax
- Surgically corrected congenital heart disease (not including catheter-based procedures)
Cardiac Related:
- Indication for permanent bradycardia pacing or cardiac resynchronization therapy
- Decompensated heart failure
- Currently on inotropic therapy
- Known history of pericardial disease, pericarditis or mediastinitis
- Risk factors associated with defibrillation testing:
- Severe aortic stenosis
- Current Intracardiac LA or LV thrombus
- Severe proximal three-vessel or left main coronary artery disease without revascularization
- Hemodynamic instability
- Unstable angina
- Recent stroke or transient ischemic attack (within the last 3 months)
- Known inadequate external defibrillation
- LVEF < 20% (most recent assessment; must be within 180 days of consent)
- LVEDD >70 mm (most recent assessment; must be within 180 days of consent)
Locations
- UC San Diego Health
not yet accepting patients
La Jolla California 92037 United States - Providence Medical Foundation
not yet accepting patients
Orange California 92868 United States - St. Bernards Heart and Vascular
accepting new patients
Jonesboro Arkansas 72401 United States
Lead Scientist at UCSD
Details
- Status
- accepting new patients at some sites,
but this study is not currently recruiting here - Start Date
- Completion Date
- (estimated)
- Sponsor
- AtaCor Medical, Inc.
- ID
- NCT07324031
- Study Type
- Interventional
- Participants
- Expecting 280 study participants
- Last Updated